Abstract

A flow injection on-line preconcentration technique, combined with graphite furnace atomic absorption spectrometry for the determination of gold in ore samples has been presented. An α-amino pyridine resin (AP) was loaded onto a microcolumn as the preconcentration reagent and a solution of 90% acetone-5% HCl-5% H 2O was used for the gold elution. The Scaled Simplex Method was employed to optimize the flow injection manifold. The selected factors were optimized by the simultaneous consideration of two responses, absorbance and column retention efficiency, with only 12 tests. A detection limit of 0.065 μg Au/l (3σ) with a relative standard deviation of 4.3%, was achieved. Ore samples containing gold have been analysed successfully using the proposed method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.