Abstract
PurposeThis study was developed to create computer software for performing the gamma index comparison between measurement and Monte Carlo (MC) simulation for the percentage depth dose (PDD) and beam off-center ratio profile (OCR). Materials and methodsThe gamma software was built in the matrix laboratory (MATLAB) software environment. The developed software was compared with ScanDosematch and Bistromath software’s gamma evaluation to assess its accuracy. A set of reference and evaluated dose distribution, which were obtained from measurement and MC simulation, was input to the software to calculate the 1D gamma index using different criteria (i.e. 3%/3 mm, 2%/3 mm, and 2%/2 mm). ResultsWe compared the two results of gamma index at 3%/3 mm, 2%/3 mm, and 2%/2 mm criteria, one calculated by the proposed software and one manually. The comparison showed high agreement between the proposed software and theoretical calculation. ConclusionsBased on the results, we concluded that our developed software has high accuracy, compared to theoretical calculation. This software could serve as a non-commercial and open-source tool for researchers and students.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.