Abstract

An open-flow helium cryostat for single-crystal X-ray diffraction experiments capable of reaching 14 K has been developed using off-the-shelf components. Solid/liquid air build-up is prevented using the transfer-line helium back-flow and a heated nozzle. The system has run for over 30 h with no frost build-up. The effectiveness of the system has been demonstrated using test data for oxalic acid, terbium vanadate and dysprosium vanadate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have