Abstract

An on-line solid phase extraction–high performance liquid chromatography–tandem mass spectrometry method for the analysis of perfluoroalkyl substances (PFASs) in water samples was developed. The optimal analytical conditions were obtained through the optimization of the extraction efficiency of on-line solid phase extraction column, sample loading rate and loading volume, and the concentration of ammonium acetate in mobile phase. Under the optimal condition, the analytical method displayed good linearity (r2 > 0.99) for 12 PFASs (C5–C14 perfluoroalkyl carboxylic acids and C6/C8 perfluoroalkyl sulfonic acids) over a concentration range of 0.5–100 ng/L. The limits of quantitation for samples were between 0.025 ng/L and 0.5 ng/L and the relative standard deviations (RSD) of five consecutive analyses were less than 10% for 1 ng/L standard solution. Satisfactory results were obtained using this analytical method for the analysis of perfluoroalkyl substances in Antarctic ice core samples. The recoveries of all perfluoroalkyl substances were in a range of 73%–117% when the samples were spiked with standards at the concentrations of 2.5 ng/L and 25 ng/L.