Abstract

Background: Hypoxemia is currently treated in hospital wards with oxygen, released continuously by “conventional” flow meters. A new type of hybrid flow meter allows to switch between on-demand and continuous mode. The aim of this observational study was to assess whether this new device reduces oxygen expenditure, is well accepted in a hospital setting and improves patient comfort during oxygen therapy. Methods: Oxygen was administered in hypoxemic patients with conventional or hybrid flow meters to maintain an oxygen saturation of ≥ 92% over a 12-week period. Every two weeks conventional and hybrid flow meters were switched. The overall oxygen delivery to the ward was continuously measured with a data logging device installed in the main oxygen pipeline and corrected for multiple confounding factors. Humidity measurements, for which a sensor placed in front of one of the nostrils, and patient questionnaires, were used to assess patient comfort during continuous and on-demand flow. Results: Overall oxygen delivery decreased by 39% when switching from continuous flow to on-demand therapy after correction for confounding factors. Continuous flows significantly decreased relative humidity more than equivalent on-demand settings and the latter tended to increase comfort. Conclusions: Hybrid flow meters cause a significant reduction in oxygen delivery in a hospital ward, which may lead to financial savings. Using the on-demand technology also lowers the dryness of the upper airways (and may increase patient comfort), while maintaining an adequate oxygenation.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call