Abstract
The discrimination of single unit activity in extracellular recordings presents a serious problem when the signal-to-noise ratio is low or when the amplitudes of interspersed spikes are similar. By exploiting spike form, the system described here performs discrimination using on-line hardware template matching. Using analog delay lines, the combined deviation of 8 input signal values from 8 stored template values is calculated simultaneously. The 8 template values are selected by adjusting 8 cursors to the desired spike trace on a CRT; the spike form discriminator (SPIFODIS) then generates a deviation function which steeply drops to zero whenever form similarity occurs, allowing for easy triggering. The performance of SPIFODIS was compared quantitatively with that of a conventional amplitude trigger in two cases: (a) when detecting a single unit with varied signal-to-noise ratios and (b) when separating double units of equal amplitude. (a) At signal-to-noise ratios between 2 and 1 the error rate for SPIFODIS was only 15–50% of that of an amplitude trigger. (b) In double-unit recordings showing only form differences, spikes are discriminated with very low error rate, while an amplitude trigger fails completely.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.