Abstract

With SiGe device technology still developing, a concomitant effort to develop accurate dopant quantification over the entire Si1−xGex (0≤x≤1) range using ultra low energy secondary ion mass spectrometry is required. Here we present a comprehensive secondary ion mass spectrometry study of the B11+ yield behaviour from a range of Si1−xGex (0≤x≤1) reference samples implanted with the same B dose. Depth profiling conditions include near normal incidence O2+ over an energy range of 0.2−1.0 keV. Quantification of the B concentration for each profile was achieved by determining the individual profile sensitivity factor. Only for EP≤300 eVwas the variation in sensitivity factor with Ge well described by a linear dependence over the whole Si1−xGex (0≤x≤1) matrix.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.