Abstract

The respiratory motion of abdominal organs is a serious obstacle in high-intensity focused ultrasound (HIFU) treatment with magnetic resonance imaging (MRI) guidance. In this study, a two-dimensional (2D) MRI-conditional motion phantom device was developed in order to evaluate HIFU protocols in synchronized and non-synchronized ablation of moving targets. The 2D phantom device simulates the respiratory motion of moving organs in both the left-right and craniocaudal directions. The device consists of MR-conditional materials which have been produced by a three-dimensional (3D) printer. The MRI compatibility of the motion phantom was tested successfully in an MRI scanner. In vitro experiments were carried out to evaluate HIFU ablation protocols that are minimally affected by target motion. It was shown that only in synchronized mode does HIFU produce thermal lesions, as tested on a gel phantom mimicking the moving target. The MRI-conditional phantom device was shown to be functional for its purpose and can be used as an evaluation tool for testing HIFU protocols for moving targets in an MRI environment. Copyright © 2015 John Wiley & Sons, Ltd.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.