Abstract
Simultaneous measurements of the circulating testosterone (TS) and dehydroepiandrosterone sulfate (DHEAS) are deemed to be helpful for the assessment of men's health. Liquid chromatography/electrospray ionization-tandem mass spectrometry (LC/ESI-MS/MS) is the most reliable methodology for this purpose; however, it has room for improvement in analysis throughput. In this study, a quadruplicate of the Girard reagents was used to develop an LC/ESI-MS/MS method capable of quantifying TS and DHEAS in four different serum samples in a single run. The four serum samples were separately pretreated, derivatized with one of four Girard reagents, and then combined. The LC/ESI-MS/MS analysis of the combined sample provided the androgen concentrations of four serum samples in parallel. The method had practical measuring ranges, in which good precision and accuracy, as well as negligible matrix effects were verified. The speed-up capability of the developed method was evaluated through the analysis of ten batches of serum samples (total 40 samples); the method saved a 60% post-pretreatment analysis time compared to the non-derivatization method for 40 samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.