Abstract

Cystatin C is considered as an alternative to the evaluation of glomerular filtration rate. In this study, we highlighted an LC-MS/MS approach for the absolute quantitation of serum cystatin C based on label-free internal standards. A tryptic peptide (ALDFAVGEYNK) was selected as the surrogate whilst analogue (ALDFAVGEYQK) served as an internal standard. After denaturation, reduction, alkylation, digestion and concentration, the target peptides were separated on an LC column and monitored under MRM. The calibration range was from 0.25 mg/L to 15 mg/L with LLOQ of 0.05 mg/L and LOD of 0.03 mg/L, respectively. The certified reference material (ERM-DA471) was determined at 5.12 mg/L with bias of 6.57%. The recovery was between 89.68% and 92.43%. The RSD of intra- and inter-assay imprecision were both <10%. Good stability was also observed. The assay also demonstrated that the quantification of native cystatin C in human serum could be achieved using label-free internal standards. The assay was robust, cheap and sensitive.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.