Abstract

The construction and operation of an isothermal microcalorimeter are described. The instrument relies on the triple point of water for the maintenance of the constant temperature envelope. This ensures a temperature stability to ±0.0001 degC. Thermistor thermometry is used and the smallest detectable temperature change in the calorimeter is 0.00012 degC corresponding to a power change of 0.5 μW. The microcalorimeter has been employed for the measurement of alpha and beta emitting radioactivities; for example, an activity of about 4 Ci of tritiated water was measured with an estimated accuracy to ±2.5%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.