Abstract

The resonant frequencies and flexural sensitivities of an atomic force microscope (AFM) assembled cantilever probe which comprises a horizontal cantilever, a vertical extension and two tips located at the free ends of the cantilever and the extension are studied. This probe makes the AFM capable of simultaneous topography at top-surface and sidewalls of microstructures especially microgears which leads to a time-saving swift scanning process. In this work, the effects of the sample surface contact stiffness and the geometrical parameters such as the ratio of the vertical extension length to the horizontal cantilever length and the distance of the vertical extension from clamped end of the horizontal cantilever on the resonant frequencies and flexural sensitivities are assessed. These geometrical effects are illustrated in some figures. The results show that the low-order vibration modes are more sensitive for low values of the contact stiffness but the situation is reversed for high values.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.