Abstract

In this paper we investigate the biostability of a series of Biolon blood sacs that were utilized in electric total artificial hearts for time periods of up to 19 weeks. A battery of experimental probes, including scanning electron microscopy (SEM) and X-ray photoelectron spectroscopy (XPS), were used to characterize the bulk and surface properties of explanted and control blood sacs. Gel permeation chromatography (GPC) experiments showed that generally there was a dramatic increase in average molecular weight at longer implantation times. However, SEM and GPC observations suggest significant deterioration of the flex regions of right blood sacs after 17 weeks of service. XPS experiments indicated appreciable silicon and hydrocarbon concentrations on blood-contacting surfaces both before and after implantation, and we speculate as to their origin.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.