Abstract
A number of problems have been reported with the use of standard length tracheostomy tubes in adult critical care patients. We measured the depth from the skin surface to the tracheal wall and the angle of the tracheal stoma during percutaneous tracheostomy. In vitro measurements were also performed on standard tracheostomy tubes. Comparison of in vivo and in vitro dimensions demonstrated that standard length tracheostomy tubes are too short for the average critical care patient. We recommend that both the stomal and intratracheal lengths should be made longer by approximately 1 cm and tubes should be redesigned to an angle of 110-120 degrees to allow optimal tracheal placement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.