We developed a personal blood pressure monitoring system for patients with hypertension or hypotension. The system can be used to measure a patient's blood pressure at home and to transmit the data automatically to a hospital database via the Internet. The accuracy of blood pressure readings using the system was assessed by comparison with readings from a standard digital sphygmomanometer in four subjects. The measurement error for the systolic readings was 1.7-2.7% and for the diastolic readings 2.7-3.2%. The system therefore appears to be a promising means of assessing blood pressure remotely.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.