In order to describe how a conduction calorimeter works, a novel model, namely the interfacial model of a conduction calorimeter, has been proposed in this paper. Based on the model, a corresponding analog thermo-analytical curve equation has also been established. For verification of its validity, thermo-analytical curves of alkaline hydrolysis of ethyl propionate and ethyl acetate have been measured at 25°C, respectively. The data calculated by this model are in reasonable agreement with those read from the thermo-analytical curves.