Abstract
A fully automated software is introduced to facilitate MR mammography (MRM) examinations and overcome subjectiveness in diagnosis using normalized maximum intensity–time ratio (nMITR) maps. These maps inherently suppress enhancements due to normal parenchyma and blood vessels that surround lesions and have natural tolerance to small field inhomogeneities and motion artifacts. The classifier embedded within the software is trained with normalized complexity and maximum nMITR of 22 lesions and tested with the features of remaining 22 lesions. Achieved diagnostic performances are 92% sensitivity, 90% specificity, 91% accuracy, 92% positive predictive value and 90% negative predictive value. DynaMammoAnalyst shortens evaluation time considerably and reduces inter and intra-observer variability by providing decision support.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.