Abstract

There is frequently a need to make laser power or energy density measurements when determining whether radiation from a laser system exceeds the Maximum Permissible Exposure (MPE) as defined in BS EN 60825. This can be achieved using standard commercially available laser power or energy measurement equipment, but some of these have shortcomings when used in this application. Calculations must be performed by the user to compare the measured value to the MPE. The measurement and calculation procedure appears complex to the nonexpert who may be performing the assessment. A novel approach is described which uses purpose designed hardware and software to simplify the process. The hardware is optimized for measuring the relatively low powers associated with MPEs. The software runs on a Psion Series 3a palmtop computer. This reduces the cost and size of the system yet allows graphical and numerical presentation of data. Data output to other software running on PCs is also possible, enabling the instrument to be used as part of a quality system. Throughout the measurement process the opportunity for user error has been minimized by the hardware and software design.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.