Abstract
This instrument includes an ionization chamber covering the whole X-ray beam in the master cone of the set, a cathode follower circuit mounted on the cone, and a trigger unit (near the control panel) based on a cold cathode relay valve. The use of special high insulation relays and electrometer valves is avoided. The indicator steps are made to represent 5 r skin dose for any treatment condition by the use of variable bias (calibrated directly in terms of, for example, applicator field size) on the trigger circuit. The calibration remains constant within ± 2% over a period of months and is sensibly independent of mains voltage variations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.