Abstract
With the invention of "open" magnetic resonance imaging (MRI) systems, access to the patient is possible during the imaging procedure. An important application of these systems is intraoperative MRI to control the extent of resection during tumor surgery. Up to now flexible surface coils wrapped around, or placed at each side of the head, were used for imaging. These flexible coils have several disadvantages such as unreliability, interindividual problems, difficult handling, poor hygienic properties, and often unsatisfactory or inhomogeneous image quality. To solve most of these problems, an MR-compatible head-holder in combination with an integrated surface coil for use in a 0.2 T C-shaped magnet was developed. Forty-eight patients with known cranial tumors underwent MRI intraoperatively. In 32 patients (67%), residual tumor was found, and additional surgical resection was performed. The integrated head-holder/coil is a safe and practical tool for intraoperative MRI, providing efficient and reliable resection control during neurosurgical procedures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.