Abstract
Electrophoretic mobility control (EMC) was used to alleviate the adverse effect of the ion-pairing agent heptafluorobutyric acid (HFBA) in the liquid chromatography–electrospray ionization mass spectrometry (LC-ESI-MS) analysis of aminoglycosides. Aminoglycosides separated by LC were directed to a connecting column before their detection via ESI. Applying an electric field across the connecting column caused the positively charged aminoglycosides to migrate toward the mass spectrometer whereas the HFBA anions remained in the junction reservoir, thus alleviating the ion suppression caused by HFBA. To accommodate the flow rate of a narrow-bore column, minimize the effect of electrophoretic mobility on separation, and facilitate the operation, an integrated EMC device with a split design was fabricated. With the proposed EMC device, the signals of aminoglycosides were enhanced by a factor of 5–85 without affecting the separation efficiency or elution order. For the analysis of aminoglycosides in bovine milk, the proposed approach demonstrates a sensitivity that is at least 10 times below the maximum residue limits set by most countries.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.