Abstract

Currently, the coronavirus disease 2019 (COVID-19) caused by the outbreak of a novel coronavirus (SARS-CoV-2) is spreading rapidly worldwide. Due to the high incidence of influenza coinciding with SARS-CoV-2, rapid detection is crucial to prevent spreading. Here, we present an integrated dual-layer microfluidic platform for specific and highly sensitive SARS-CoV-2, influenza viruses A (FluA) H1N1, H3N2, and influenza virus B (FluB) simultaneous detection. The platform includes a dual microchip (Dμchip) and a portable detection device for real-time fluorescence detection, temperature control and online communication. The Reverse Transcription Loop-mediated Isothermal Amplification (RT-LAMP) and Cas12a cleavage were performed on the Dμchip. The limit of detection (LoD) of the Dμchip assay was 10 copies for SARS-CoV-2, FluA H1N1, H3N2, and FluB RNAs. The Dμchip assay yielded no cross-reactivity against other coronaviruses, so it was suitable for the screening of multiple viruses. Moreover, the positive percentage agreement (PPA) and negative percentage agreement (NPA) of the assay were 97.9% and 100%, respectively, in 75 clinical samples compared to data from RT-PCR-based assays. Furthermore, the assay allowed the detection SARS-CoV-2 and influenza viruses in spiked samples. Overall, the present platform would provide a rapid method for the screening of multiple viruses in hospital emergency, community and primary care settings and facilitate the remote diagnosis and outbreak control of the COVID-19.