Abstract

The Absolute Radiometer for Cosmology, Astrophysics, and Diffuse Emission (ARCADE) is a balloon-borne instrument to measure the temperature of the cosmic microwave background at centimeter wavelengths. ARCADE uses narrow-band cryogenic radiometers to compare the sky to an external full-aperture calibrator. To minimize potential sources of systematic error, ARCADE uses a novel open-aperture design which maintains the antennas and calibrator at temperatures near 3 K at the mouth of an open bucket Dewar, without windows or other warm objects between the antennas and the sky. We discuss the design and performance of the ARCADE instrument from its 2001 and 2003 flights.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.