A new semiautomated instrument is described which quantifies the number of bacteria in potable waters within 3 min, providing a permanent colorimetric record of the results. The bacteria detection device can measure as few as 100 CFU/ml in potable waters. In brief, a 100- to 1,000-ml sample of tap water is drawn through a large surface, customized filter housed in the device, and bacteria, rust, and humic acid in the water are concentrated thereon. A reducing agent is used to remove the rust and humic acid from the filter. The filter is inverted and backflushed to elute the bacteria which are collected and reconcentrated onto a 7-mm-diameter filter surface. The reconcentrated bacteria are stained, and the filter fibers are preferentially decolorized without removing the dye from the bacteria. The color intensity of the filter surface is compared with a color guide to determine the amount of bacteria in the test water.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.