Abstract

To measure the unattached fraction of radon progeny, a new type of portable, low cost,reusable integrating monitor with etched track detector (CR-39) was developed. This monitorsimultaneously gives the average equilibrium-equivalent concentrations of unattached 222Rn progeny (EECRnu), attached 222Rn progeny (EECRna) andalso 220Rn progeny (EECTn). It was calculated, according to the detection efficiencies which werederived by the Monte Carlo method, that the lower detection limits forEECRnu,EECRna andEECTn are 0.03 Bq m−3, 0.12 Bq m−3 and0.07 Bq m−3, respectively for 24 h integrating sampling at a flow rate of 2.92 l min−1. The validity of the new monitor was confirmed through a series of intercomparisonexperiments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.