Abstract
Abdominal compartment syndrome can be treated through decompressive surgery if intraabdominal hypertension (IAH) can be detected in time. Treatment delays due to manual, conventional intravesical pressure (IVP) monitoring using a Foley catheter have been reported. In this work, we present an innovative gastrointestinal intraluminal pressure (GIP) measurement-based method to monitor and improve pressure-guided relief of intraabdominal pressure (IAP). A novel algorithm for detecting IAH in the gastrointestinal tract of a live porcine model is reported. A wireless pressure-sensing capsule (10 × 13 mm) was developed for absolute measurement. The IAP was estimated during artificial pneumoperitoneum. The pressure waveform-based measurements indicated that the wireless pressure sensor could be used to predict IAP. To enhance GIP monitoring for predicting IAH, the proposed continuous ingestible wireless electronics-based pressure waveform measurement device can be used as a complement to existing modalities. The use of the proposed pressure measurement and communication technology can help provide valuable data for digital health platforms.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.