Abstract

Electret microphones were evaluated for use in an experiment to investigate the surface pressures of a wing immersed in a propeller slipstream. Calibration of a typical electret microphone over a range of 50 to 6000 hz showed that at each individual frequency there is less than 1.1% error in linearity for low pressures, but some variation of sensitivity occurs over the range of frequencies tested. This variation of sensitivity with frequency can be corrected using standard signal processing techniques.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.