Abstract

We have developed a small, light, passive, inexpensive, datalogging particle monitor called the "UCB" (University of California Berkeley particle monitor). Following previously published laboratory assessments, we present here results of tests of its performance in field settings at high particle concentrations. We demonstrate the mass sensitivity of the UCB in relation to gravimetric filter-based PM(2.5) mass estimates as well as commercial light-scattering instruments co-located in field chamber tests and in kitchens of wood-burning households. The coefficient of variation of the unadjusted UCB mass response in relation to gravimetric estimates was 15%. Although requiring adjustment for differences in sensitivity, inter-monitor performance was consistently high (r(2) > 0.99). Moreover, the UCB can consistently estimate PM(2.5) mass concentrations in wood-burning kitchens (Pearson r(2) = 0.89; N = 99), with good agreement between duplicate measures (Pearson r(2) = 0.94; N = 88). In addition, with appropriate cleaning of the sensing chamber, UCB mass sensitivity does not decrease with time when used intensively in open woodfire kitchens, demonstrating the significant potential of this monitor.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call