Abstract

The osteoconductive properties of hydroxyapatite (HA)-coated titanium implants are well documented, but eventual coating degradation may result in HA particles adjacent to the substrate, and if the particles were to migrate into the joint space then accelerated polyethylene wear might be expected. As an indirect indication of third-body wear, the authors used laser interference microscopy to measure and compare surface roughness on modular heads from 15 clinically retrieved HA-coated femoral components, with heads from 15 retrieved uncemented (porous) and 15 cemented implants. The results showed increased median surface roughness over initial manufacturer specifications in all groups, but the cobalt-chrome heads from the HA-coated group showed significantly less surface roughness and less deep scratches than the heads from either the porous or uncemented group. Three-body wear appears to be a common problem, but evidence available suggests that it is no more of a problem with HA-coated devices than porous or cemented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.