Abstract

The accuracy of many apex locators is affected by electrolytes including sodium hypochlorite. According to the manufacturer a new device, the ENDEX, overcomes this problem. The purpose of this study was to assess the accuracy of the ENDEX in comparison with a traditional device, the RCM Mark II. The lengths of 61 canals with various contents (vital pulp, necrotic pulp, pus/exudate, sodium hypochlorite, and water) were determined in vivo. Files were cemented within the canals at lengths determined by the ENDEX. The teeth were then radiographed and extracted, and the distance between the file tip and apical foramen was recorded. Derived readings for the RCM Mark II were compared with the actual ENDEX reading and corresponding radiograph. The results indicated that most of the file tips that were at the radiographic apex actually extended through the apical foramen, and canal content had a varying effect on each device. The overall accuracy of the ENDEX (71.7%) was higher than that of the RCM Mark II (43.5%) within 0.5 mm of the apical foramen. The ENDEX, unlike the RCM Mark II, proved accurate in the presence of sodium hypochlorite.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.