Abstract

We have developed an in vivo pressure monitoring system to study the phenomenon of reverse ultrafiltration of dialysate during high-flux and high-efficiency dialysis. Under the usual operating conditions of either type of dialysis, driving pressures existed for reverse ultrafiltration of dialysate into the venous end of the blood compartment. Whether or not reverse ultrafiltration could be abolished at higher ultrafiltration rates was dialyzer-dependent, being least with high-efficiency dialysis. In addition, the degree of reverse ultrafiltration was affected by patients hematocrit, dialyzer inlet and outlet oncotic pressures, and the rate and direction of dialysate flow.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.