Abstract
CO2 lasers have been generally used in oral soft-tissue surgery. Although an increase in power enhances the depth of ablation, the specific correlation between units of power and ablative depth has not been stated. This study aimed to explore ablative depth and width affected by a power unit of 10,600-nm CO2 laser continuous wave at 3 to 10 W in tissue blocks of the swine tongue. The 112 samples were randomly allocated into eight groups according to the power settings. The depth and width of the incision were measured by using the ImageJ program. The 3 W group showed the minimal depth of ablation [0.527 mm (0.474 and 0.817)] and width of ablation [0.147 mm (0.110 to 0.184)]. The maximal depth [3.750 mm (3.362 and 4.118)] and width [0.700 mm (0.541 to 0.860)] were in the 9 W group. The correlation between the ablative depth and power unit was 0.81 (p < 0.001). Based on the regression equation (p < 0.001), the ablative depth (mm) = (the power unit of laser in W × 0.491) − 0.731. In conclusion, the ablative depth of the CO2 laser on soft tissue was strongly correlated to power units enabling the prediction equation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.