Abstract
The present study aims at evaluating the efficacy of commercially available RITA’s StarBurst® XL multi-tine electrode in attaining large size coagulation volumes (≥3 cm in diameter) during radiofrequency ablation (RFA) application. In vitro studies have been conducted on the cylindrical shaped polyacrylamide based tissue-mimicking phantom gels utilizing different active lengths of the multi-tine electrode, viz., 2 cm, 3 cm, 4 cm and 5 cm. A temperature-controlled RFA has been performed at a target tip temperature of 95 °C for 5 min. The variations in the power supply, the target tip temperature and the size of coagulation volume have been reported for different active lengths of the multi-tine electrode. The study revealed that the increase in active length of the multi-tine electrode results in more energy deposition and consequent rise in the coagulation volume during RFA procedure. Further, a simplified novel statistical correlation between the coagulation volume and active length of the multi-tine electrode has been proposed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.