Abstract

The objective of this study was to investigate venous collapse (VC) related to venous drainage during the use of an extracorporeal life support circuit. A mock circulation was built containing a centrifugal pump and a collapsible vena cava model to simulate VC under controlled conditions. Animal experiments were performed for in vivo verification. Changing pump speed had a different impact on flow during a collapsed and a distended caval vein in both models. Flow measurement in combination with pump speed interventions allows for the detection and quantitative assessment of the degree of VC. Additionally, it was verified that a quick reversal of a VC situation could be achieved by a two-step pump speed intervention, which also proved to be more effective than a straightforward decrease in pump speed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.