Abstract
The introduction of IMRT requires users to confirm that the isodose distributions and relative doses calculated by their planning system match the doses delivered by their linear accelerators. To this end the commercially available software, VeriSoft (PTW-Freiburg, Germany) was trialled to determine if the tools and functions it offered would be of benefit to this process. The CMS XiO (Computerized Medical System, St. Louis, MO) treatment planning system was used to generate IMRT plans that were delivered with an upgraded Elekta SL15 linac. Kodak EDR2 film sandwiched in RW3 solid water (PTW-Freiburg, Germany) was used to measure the IMRT fields delivered with 6 MV photons. The isodose and profiles measured with the film generally agreed to within +/- 3% or +/- 3 mm with the planned doses, in some regions (outside the field) the match fell to within +/- 5%. The isodose distributions of the planning system and the film could be compared on screen, allowing for electronic records of the comparison to be kept if desired. The features of this software would be of benefit to an IMRT QA program.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Australasian Physics & Engineering Sciences in Medicine
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.