In this study we have demonstrated that the line start technique can be used with the Horiba centrifugal particle size distribution analyzer to obtain high-quality particle size and size distribution data over a range of sizes in the micrometer and submicrometer range. The line start technique yields much better accuracy and reproducibility than the uniform distribution technique previously used with this machine. Reasons for this are enumerated in the text. Equipment needed to perform the analysis of particle size with the Horiba machine using the line start technique is now available from the Horiba subsidiary located in North America.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.