Abstract
Aiming at the scatter problem in Cone Beam Computed Tomography (CBCT) system, an improved scatter correction method based on Beam Attenuation Array (BAA) was proposed. In this method, BAA and filter plate are used as the scatter detection devices to scan 3 times and acquire 3 sets of projection images. The scatter values on the projection centers of the small ball array of BAA are calculated by using of the characteristic of a part of the ray attenuation caused by BAA, and then, the whole scatter intensity distribution covered the projection image, that is, the scatter image, is obtained by fitting these discrete scatter values on the projection centers of the small ball array. To complete the scatter correction, the final projection image is got by subtracting the corresponding scatter image from the original projection image. The experimental result shows that the improved method is practical and effective to correct the scatter in CBCT imaging systems, and the slice image quality is enhanced.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.