Abstract

A new microfluidic device to enhance the enrichment factor in miniaturized systems is proposed. The microfluidic system was design for liquid phase microextractions, and it was applied to the simultaneous extraction of acidic compounds of a wide range of polarity (0.5 < log P < 3). The device operated under stagnant acceptor phase conditions and all the operational parameters involved were optimized. Tributyl phosphate was found to be a new highly efficient supported liquid membrane to simultaneously extract analytes of very different polarities. The optimal donor and acceptor phase were pH 2 and pH 13, respectively. The donor flow rate and the extraction time were investigated simultaneously, offering great versatility with high enrichment factors (EFs). Limits of quantitation were within 0.02 and 0.09 µg mL−1 for all compounds at 10 µL min−1 as donor flow rate and 20-min extractions, offering EFs between 11 and 18 with only 200-µL sample volume consumption. The method was successfully applied to human urine samples, observing recoveries between 47 and 90% for all compounds. This new proposed microfluidic system increases the wide range of applications, especially when the analytes are present in lower concentrations in the sample.Graphical