Abstract
The capability of ICP-MS equipped with the high matrix introduction system (HMI) for accurate analysis of lanthanoids in environmental samples was investigated. Compared to the conventional operation, the amounts of oxide and hydroxide molecular species formed in the plasma were reduced by up to 5 times. The relative yields of oxides did not exceed 0.02% for BaO + species and were as low as 0.3% for lanthanoids with the highest oxide-formation rates (LaO +, CeO +, PrO + and NdO +). Hydroxide formation was less than 0.02% when HMI system was used. In addition, two digestion procedures were evaluated by the analysis of standard reference materials (SRMs) of different matrices. The digestion efficiency and limits of detection (LOD) were improved when samples were digested with a mixture of HNO 3/H 2O 2/HCl/HF, and HF was removed by evaporation in presence of concentrated HCl. Using this procedure and HMI-ICP-MS analysis, LOD ranged from 0.1 μg kg −1 to 6 μg kg −1. Recoveries ranged from 85 to 115% for La to Ho and from 75 to 85% for the other lanthanoids. Relative standard deviations for replicate analysis of SRMs were less than 10%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.