Abstract
An improved lead resistance compensation technique for use with three-wire resistance temperature detectors is presented. The present practice of using the three-wire concept in an unbalanced bridge to provide signal to direct reading instruments produces nearly 1% error for a lead resistance as low as 1 /spl Omega/. In contrast, the new method limits the error to below 0.1% with much higher lead wire resistances.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.