Abstract
An improved high performance liquid chromatography (HPLC) method for the analysis of the main organic compounds in musts and wines is presented. A column packed with hydrogen sulfonated divinyl benzene-styrene copolymer and two detectors connected in series were used (UV at 210 nm and RI, respectively). The addition of acetonitrile (6%) to the mobile phase (0.045N H2SO4) allowed the simultaneous separation and quantification of several organic acids, glucose, fructose, glycerol, and ethanol. Direct injection and sample clean-up with a SAX cartridge was tested and compared. The SAX fractionation gave satisfactory results, however the direct injection of diluted wine (1:20) provided the best precision (CV ≤ 2.1%) and accuracy of analysis. Statistical analysis (paired t-test) disclosed significant differences only for glucose, fructose, α-ketoglutaric, and pyruvic acid. The role of column temperature for the analysis of vitamin C (ascorbic acid) was also investigated. Vitamin C underwent thermal degradation during analysis with column temperature ≥ 30°C.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Liquid Chromatography & Related Technologies
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.