Abstract

In this paper we present an improved method for the selective and sensitive determination of oxalate in different matrices such as urine, plasma, and food. The method uses ion chromatography for the separation of anions. To overcome problems with interfering matrix-anions, colourings, and macromolecules, we used an inline enzyme-reactor (ER) containing immobilised oxalate oxidase, which converts oxalate to hydrogen peroxide. Hydrogen peroxide was analysed with high sensitivity by amperometric detection. The determination limit for the HPLC-ER method was 1.5 μol/l, the mean recovery in urine was 102%. The evaluation in a urinary matrix achieved CV values from 2.2% to 6.7% for the within-run precision and CV values from 3.7% to 8.6% for the between-batch precision. The results of the new method were statistically equivalent to those obtained by enzymatic kits. We present first results of the HPLC-ER method, when applied to body fluids and food analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.