Abstract

A single and sensitive spectrophotometric method for chloride ions determination based on a commuted flow system with a 100 cm optical path flow cell and a solid-phase reactor containing immobilized silver chloranilate was proposed. This procedure exploited the AgCl formation in the solid-phase reactor leading the chloranilate ions, monitored spectrophotometrically at 530 nm. The analytical signals were 75-fold higher and the sensitivity was 12-fold than that achieved with a 1 cm flow cell, allowed a chloride determination in the 0.5–100 mg l −1 range. The R.S.D. was 1.1% ( n = 20) with a sample throughput of 80 h −1 and a waste generated of ca. 100 ng of chloranilate ions per determination. Four samples of natural waters from São Carlos and Araraquara cities were evaluated using the proposed method. Results agreed with the obtained by a reference method at the 95% confidence level.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.