Abstract

In recent years, efforts have been made on the information extraction method in x-ray grating-based phase contrast imaging. Among them, the reverse projection method proposed by P.P. Zhu in 2010 features as a fast, low-dose method to extract information without stepping the gratings. It enables to extend the phase contrast imaging system to the vivo studies. However, it has strict requirement on the system stability and the source stability. In this manuscript, an improved acquisition scheme has been proposed using a staggered grating making the reverse projection method more efficient with a lower requirement of the system and source stability.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.