Abstract
A closed loop system of diabetes control would minimize hyperglycemia and hypoglycemia. We therefore implanted and tested a subcutaneous amperometric glucose sensor array in alloxan-diabetic rats. Each array employed four sensing units, the outputs of which were processed in real time to yield a unified signal. We utilized a gain-scheduled insulin control algorithm which rapidly reduced insulin delivery as glucose concentration declined. Such a system was generally effective in controlling glycemia and the degree of lag between blood glucose and the sensor signal was usually 3-8 min. After prolonged implantation, this lag was sometimes longer, which led to impairment of sensor accuracy. Using a prospective two-point calibration method, sensor accuracy and closed loop control were good. A revised algorithm yielded better glycemic control than the initial algorithm did. Future research needs to further improve calibration methods and reduce foreign body fibrosis in order to avoid a time-related increase in lag duration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.