Abstract
In this paper, we describe an in situ sensor for estimating bacterial concentration in a lab-scale fermentor using impedance spectroscopy. We constructed the impedimetric biosensor using a gold-coated silicon wafer, PDMS polymer, and a borosilicate glass tube. An advantage of using these materials was that we could make a plug-type, disposable electrode. We compared the impedimetric differences between bacterial cells, culture media, and metabolic by-products. We measured real-time impedance changes of an E. coli solution at 0.01 MHz frequency of applied AC for 13 h, and correlated them with the bacteria concentration as estimated by a hemocytometer. In conclusion, impedance increased and reactance decreased during the measurement period, as the concentration of E. coli increased. The correlation between the E. coli concentration and the impedimetric data was linear.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.