Abstract

Ochratoxin A (OTA) is a mycotoxin that can cause health risks to human/animal health. Contamination by OTA can occur in various foods and agricultural products, so sensitive and rapid detection of OTA is crucial. We describe a simple and sensitive fluorescence immunoassay for OTA using tetramethylrhodamine (TMR)-labeled OTA as a fluorescent probe. We conjugated tetramethylrhodamine to OTA through a covalent reaction, and obtained three TMR-OTA isomer probes after purification by high-performance liquid chromatography. All of the fluorescent probes showed high binding affinity to the anti-OTA antibody. Binding of the TMR-OTA probe to the antibody induced strong fluorescence of TMR-OTA due to the possible change in the local environment of TMR caused by affinity binding. In the presence of OTA, the OTA target competitively displaced the bound TMR-OTA probe from the antibody, causing a decrease in fluorescence. Measuring the change in fluorescence enabled rapid detection of OTA. This method was selective and allowed the detection of 1 nM OTA, showing potential for rapid OTA analysis in applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.