Abstract
A simple system, which allows an ion source gas bottle (located in the terminal of the accelerator) to be pressurized through copper and polypropylene tubing from an external gas cylinder, is described. Since it is not necessary to remove the insulating gas from the accelerator tank, a complete change in accelerated ions can be made in 30 min or less. Electrical breakdowns in the polypropylene tubing (with subsequent rupture) is prevented by backfilling with insulating gas.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.