Abstract

The NCPA has developed an external calibrator system which is nearing maturity. This calibrator is attached as a replacement sensor lid for the Hyperion sensor, and injects the signal into the back volume of the Hyperion sensor. When the external calibrator is mated to the sensor, the integrated sensor-external calibrator package behaves as a self-calibrating sensor. The external calibrator allows the Hyperion to operate nominally and without significant change in response with the external calibrator installed. The calibrator is driven by an external signal generated by the digitizer (e.g., the CAL signal on a GEOTECH). This system is capable of producing signals with amplitudes greater than 20-Pa in the operational environment between 0.01 and 10 Hz. We report here on the performance metrics (frequency flatness, level linearity, etc.) as well as summarize the theory of operation of the device.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.