Abstract

Only a few methods exist to observe, identify, and localize defects in SiC devices. These defects are a major limit for device performance and reliability. Presented is an improved experimental setup to investigate deep level defects using electrically detected magnetic resonance (EDMR). The method applied in this study exploits the simultaneous in-situ electron spin resonance (ESR) measurement of a standard sample (DPPH) to calibrate the magnetic field. The functionality is shown by comparing the data of an ion implanted SiC diode to results from a recent study . The in-situ B-field calibration is found to increase the accuracy of EDMR measurements by a factor of 2.5.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.